Analysis of 12 kinds of antiseptics in cosmetics using Unisol C18(2)

Application Introduction

Favorite

Introduction

The application note demonstrated the analysis of 12 kinds of antiseptics in cosmetics using Unisol C18(2) column.

                                                                 Table 1 Information of analytes

Analytes

Formula

Molecular Formula

Molecular Weight

CAS

 Metatin GT

 

    C4H5Cl2NOS

186.0596

26172-55-4

Bronopol

C3H6BrNO4

199.988

52-51-7

2-Methyl-4-Isothiazoline-3-one

 C4H5NOS

115.15

2682-20-4

Benzyl alcohol

C7H8O

108.1378

100-51-6

2-Phenoxyethanol

C8H10O2

138.1638

122-99-6

Methyl 4-hydroxybenzoate

C8H8O3

151.1399

99-76-3

Benzoic acid

C7H6O2

122.1224

65-85-0

Ethyl 4-hydroxybenzoate

C9H10O3

166.1739

120-47-8

Isopropyl paraben

C10H12O3

180.2005

4191-73-5

Propyl 4-hydroxybenzoate

C10H12O3

180.2005

94-13-3

isobutyl 4-hydroxybenzoate

C11H14O3

194.2271

4247-02-3

Butyl paraben

C11H14O3

194.2271

94-26-8

Experimental

Material and Reagents

Watsons water; HPLC purity of methanol and acetonitrile; analytical pure of sodium dihydrogen phosphate and phosphoric acid;

Standard solution of 12 antiseptics mix: provided by customer;

Hand sanitizer: bought from supermarket;

Blank sample: Transfer 1 g hand sanitizer into a 10 mL conical flask with cover, add 10 mL methanol and sonicate for 15 min, filter the extract through 0.45 mm filter;

Blend of blank sample and standard solution: Take 30 mL blank sample solution and 70 mL standard solution mixed respectively, mix well.

Instrumentation

Column: Unisol C18(2), 5 µm, 110 Å, 4.6 × 250 mm;

Mobile phase: A: acetonitrile; B: 0.05 mol/L of sodium dihydrogen phosphate (adjust the pH to 3.0 with phosphoric acid); C: methanol;

Detection: 254 nm;

Flow rate: 0.8 mL/min;

Column temperature: 35 °C;

Injection: 5 mL;

                                                                          Table 2 Gradient

Time(min)

A%

B%

C%

0

5

90

5

5

20

70

10

20

20

40

40

30

30

40

30

30.1

5

90

5

40

5

90

5

Results and Discussion

We developed a robust analytical method for quantifying 12 kinds of antiseptics in cosmetics. Excellent chromatographic separation was obtained using Unisol C18(2) for each analyte.

                Table 3 Performance of Unisol C18(2) in analysis of standard solution of 12 antiseptics mixed

Analytes

Retention   Time

Number   of Theoretical Plates

Tailing   Factor

Resolution

Notes

 Metatin   GT

6.555

31976

1.09

Figure 1

Bronopol

8.936

63026

1.03

16.45

2-Methyl-4-Isothiazoline-3-one

10.842

99934

1.10

13.61

Benzyl   alcohol

13.653

82319

1.05

17.14

2-Phenoxyethanol

15.952

92684

1.04

11.48

Methyl   4-hydroxybenzoate

16.421

94142

1.08

2.21

Benzoic   acid

17.030

104788

1.01

2.86

Ethyl   4-hydroxybenzoate

21.546

150562

1.04

20.85

Isopropyl   paraben

25.404

182440

1.03

16.75

Propyl   4-hydroxybenzoate

26.252

165986

1.04

3.41

isobutyl   4-hydroxybenzoate

31.813

125141

1.04

17.99

Butyl   paraben

32.454

123021

1.04

1.75

                                             Table 4 Performance of Unisol C18(2) in analysis of blank sample

Analytes

Retention Time

Number of   Theoretical Plates

Tailing Factor

Resolution

Notes

Metatin GT

6.554

33909

0.99

Figure   2

2-Phenoxyethanol

15.948

93667

1.04

53.46

      Figure 1 Chromatography of 12 antiseptics standard solution mix


                      Figure 2 Chromatography of blank sample

Figure 3 Chromatography overlapped of standard solution and blank sample

Figure 4 Chromatography of blank sample and standard solution blend

Conclusion

We developed a robust analytical method for quantifying 12 kinds of antiseptic in cosmetics. Excellent chromatographic separation was obtained using Unisol C18(2) for each analyte. The result described that two kinds of antiseptics were pinpointed in the hand sanitizer samples and there was no interference of matrix for the separation of antiseptics. Unisol C18(2) delivered robustness performance and thoroughly meet the demand of safety and regulatory testing.

Ordering Information

Products

Specification

Cat.No

Unisol   C18(2)

5 μm,110 Å, 4.6×250 mm

VA952505-2

1.5 mL vials

Screw neck vials,   12 × 32 mm

AV1001-6

Caps and Septa

Screw neck cap,   center hole; red silicone/ white PTEE septa, slitted

AV2200-0

Syringe   Filter (Nylon)

Monofilm,   13 mm, 0.45 μm

AS021345-T

Disposable   Needle-Free injection systems

2 mL, 100/pk

LZSQ-2ML







 

Download
Material Name Material Type Date
Modify product number success!
Total Price:

Close

Email To Sombody

X
Email
Subject
Content

Login

X
Email address
Password
Register Forgotten your password?